1.00 min: left click Workflow Sequencer
Sequence: For acquisition controlling the TOF
Processing: post processing. Eg integration, deconvolution
History: past runs
2.00 min. Left click on Sequence tab
2.10 min creating new sequence
Choose <new project> can create and named new project
Left click this can choose the path where is it being saved.
Then name the sequence file. Can have multiple sequences in the same project
3.50 to delete the sequence right click can delete
4.13 min: performing AO (auto-optimisation: we call it tuning in shimadzu)
Left click on the newly created project AO
For file name avoid using scripting symbols like µ # // this sort, will cause bug
Comment is fine as it is
{d} this is the date
{n} number auto increment, if same file name first run will be 1, and so on and so fro
If same file name, computer actually will auto put a time stamp behind
6.29 min. Double left click Global Method
Choose template method. Template method here are hard coded. Cannot be over written.
Choose auto-optimise BenchTOF. It will load onto the Global Method.
Right click view method
Pre-run pause: should not need to touch. don’t touch. Slight delay before TOF starts collecting data.
Filament voltage also don’t touch just leave it as 1.6 V. For brand new system, should not need to change this above 1.6V
Define transfer line and ion source temperature, make sure they are the same as the acquisition method.
Mode selection drop down option
We will do standard AO (at 70eV) for this exercise. Can use this to just do 14ev if we just want to tune soft ionization.
We can click on the pad lock and unlock it. Save it as a new name
Choose Yes
Rename the file
Now we can edit the parameters in the sequence
This is useful for method development can run sequence of multiple setting. Rather than save eg 10 different method just because of slight difference in parameters.
If select TI, we will have one more mode of auto-optimisation
Lower ionization, drop in sensitivity. If struggle to see try -16, but if have slightly higher in abundance can try -12.
The library trainer passed us is at -14.
14min: if a warning popup when we save the method due to change in parameters of the source. This is normal. No need be alarm. Will not see such warning if the method save the parameters is same as current setting.
14.45: the tune files are saved in the background configuration. If you run a acquisition method that it does not have the tune file for it, the software will choose the tune file closest to it.
Eg you did not tune at -14 eV but you perform acquisition at -14 eV. You have a tune file at -16 eV. It will use that
A warning popup will appear.
16.00: tune file will keep overwrite the last one for the similar eV
To view old tune file info
Right click on BenchTOF, choose generate report.
Can generate report according to mode: Standard AO or Tandem AO,
Ionization energy:
Or filament current
18.00 min starting the auto-optimization tuning
Starting Click play.
It will perform short validation to make sure file path is ok. If make an external drive and unplugged will cause issue.
Instrument will hang and stopped if drive not found.
After clicking play a popup will appear. Ignore them all when doing a tune.
The repeat features is useful to keep repeating that line. Flush contamination until its cleared.
23.02 min during auto tuning.
First thing it does is make sure getting ionisation
Then check for electronic noise on detector by scanning the whole mass range. This process is known as DarkCounts
If very high Darkcount means something wrong with detector maybe got dust on MCP.
The !!! marks are signs the detectors are overloaded (very normal during AO, ignore)
It also check for SigmaNoise. That is the electronic noise
26.55 min Acceptance criteria
low eV (14 or 16)
70 eV
For 70 eV mass resolution above 1100 = passed else fail
For 14 or 16 eV mass resolution above 800 = passed else fail
Ion intensity (70 eV) 2000000 passed (2 million)
Ion intensity (low eV) 500 000 (0.5 million)
29.40 min: generating of report note the value here have to divide by 40 as its using older scripting for older instruments of Sepsolve
Right click on
Choose Generate Report. On the new popup choose the setting required
Values here divide by 40 will be same as what we got in the sequence view.